Extra Column Dispersion – Low Dispersion Gradient Elution Apparatus > Page 51

This usually means reducing the displacement volume of the piston from, perhaps 250 ml to 25 ml or less. The output from the two pumps is passed to a special low volume mixing T, a diagram of which is shown in figure 25.

Figure 26. Block Diagram of Gradient Elution Apparatus for Small Bore Columns

Figure 26. Block Diagram of Gradient Elution Apparatus for Small Bore Columns