Column Packing
There are a wide range of different packings (both in type and particle diameter) that can be used in both GC and LC. However, there are two basic types of packing, supports that carry the stationary phase (e.g. Celite or calcined Celite used mostly in GC) and materials that act as the stationary phase itself (silica gel and bonded phases used mostly in LC). Celite is a very inactive form of almost pure silica that comprise the skeletons of diatoms deposited in seas millions of years past. To reduce its residual activity, which is largely due to surface hydroxyl groups and heavy metal ions, it is washed with acid, calcined and finally treated with hexamethyl disilazane to remove the last remaining surface hydroxyl groups. The material is screened to particle diameters ranging between 150 m and 200 m, the closer the cut of particle size, the easier the column is to pack and the greater the efficiency that will be obtained. The support is coated with the stationary phase by adding a solution of the stationary phase in a volatile solvent to a weighed amount of support and evaporating the solvent in a rotary evaporator. A dry packing process is usually employed for packing GC columns. Stationary phase loading for GC columns range from 3-4 % w/w to about 15 % w/w. Silica gel and bonded phases for use in LC are prepared in a range of particle diameters, particles 3 m in diameter are used for packing short (3 cm) columns, particles 5 m in diameter are employed for packing columns 5 to 10 cm long and particle diameters of 10 m for packing 20 cm long columns. Particle diameters of 20 m can be used to pack columns up to 1 m in length (the packing procedure, however, can be very difficult to pack). LC columns are slurry packed under high pressures.
Author: RPW Scott
Book:Preparative Chromatography
Section:Preparative Packing-Preparative-Columns LC-Columns
When slurry
packing preparative LC columns, care must be taken not to operate at pressures
in excess of the bursting strength of the tube used for the column. As the
column diameter increases, the maximum permissible pressure rapidly falls
unless extremely thick walled tubing is employed. The safe maximum pressure for
any tube can usually be obtained from the tube or column supplier. A slurry is
made of the packing (110% of that needed to fill the column) and placed in the
packing reservoir. The reservoir is rapidly connected to the pump (which must
have both an adequate delivery rate and an adequate operating pressure - the
delivery rate will depend on the column diameter and the applied pressure on
the wall thickness and the column length). The pump is started and the column
exit valve opened and the flow continued until it has been reduced to a
constant value. The flow is then arrested and some packing will remain in the
reservoir
Preparative Packing-Preparative-Columns LC-Columns
Author: RPW Scott
Book:Preparative Chromatography
Section:Preparative Packing-Preparative-Columns LC-Columns
The pre-column
is then carefully removed from the actual column and the top flange connected.
The pre-column ensures that the top of the packing will be packed to a similar
density to the bulk and will not be 'loose' and more porous and permeable.
For larger
columns the apparatus on the left of figure 14 can be used. This procedure was
developed by Filipi (9) for small bore GC columns but works equally well for
large preparative columns. Sufficient packing is placed in the top reservoir to
pack about 110% of the column volume, so that after packing the reservoir still
contains a significant amount of packing. The reservoir is then connected to a
reducing valve and gas cylinder (tank). The valve at the column exit is closed
during this period. The pressure in the system is very slowly brought up to a
pressure of about 50 p.s.i. (the optimum pressure will depend to some extent on
the particle size of the packing and the column length and may
Preparative Packing-Preparative-Columns LC-Columns
Author: RPW Scott
Book:Preparative Chromatography
Section:Preparative Packing-Preparative-Columns Axial-Compression
hydraulic
jack, can be moved throughout the entire length of the column. It contains a porous
frit at the top and a channel for passage for the mobile phase through its
center. The channel is connected to the detector and thence to the fraction
collector. To pack the column, the piston is withdrawn to bottom of the column
and the column filled with the measured mass of packing. The amount of packing
used will determine the effective length of the column. The hydraulic jack is
activated and the packing compressed to the maximum pressure which is
determined by the mechanical strength of the packing. If the packing is crushed
by the use of too great a pressure, many fines will be produced, which cause
the permeability to be extremely low and seriously reduce the column
efficiency. The pressure to the column is maintained continuously and during
each separation. In addition to stabilizing the bed when packed, the system
allows the rapid packing and unpacking of media which can range from
Preparative Packing-Preparative-Columns Axial-Compression
Author: RPW Scott
Book:Gas Chromatography
Section:YES GC-Columns Packing
13.
Figure 13 An Example of a
Column Packing Apparatus
The packing is
placed in a reservoir attached to a gas supply that forces the packing through
the column. The column exit is connected to a vacuum pump. A wad of quartz wool
is placed at the end of the column, constrained by a small restriction, that
prevents the wad from being sucked into the pump. The vacuum and gas flow are
turned on simultaneously and the packing is swept rapidly through the column.
This causes the material to be slightly compacted along the total length of the
column and has been shown to produce well-packed columns. The procedure is a
little tedious and the success rate is sometimes less than 90%. In addition,
the process does not lend itself to automation.
The
difficulties involved preparing packed columns have also contributed to the
preferential popularity of the open tubular columns. The production of capillary columns can be
largely automated and
YES GC-Columns Packing
Author: RPW Scott
Book:Preparative Chromatography
Section:Preparative Introduction
by increasing the dimensions of the column
both in GC and in HPLC. However, this approach has distinct limitations.
If the column
radius is increased, unless special packing techniques are employed, the
packing procedure becomes inefficient and the packing itself unstable. In
addition to maintain the optimum mobile phase velocity, the flow rate will need
to be substantially increased and the consumption of mobile phase will
eventually become economically impractical. Conversely, if the column length is
increased, then the impedance to flow will become greater leading to high
column pressures. If large column radii are employed, then the mechanical
strength of the column system will limit the maximum permissible pressure. Consequently, lengthening the column will eventually require the particle
diameter to be increased to provide adequate permeability. Increased particle
diameter will, in turn, reduce the column efficiency, which may impair the
resolution of the compounds of
Preparative Introduction
Author: RPW Scott
Book:Gas Chromatography
Section:YES GC-Columns Packing
the
support and an accurate value for the stationary phase loading.
Column Packing
Short columns
are usually straight and can be packed vertically. The packing is added, about
0.5 ml at a time, and the column tapped until the packing had settled. Another
portion of packing is then added and the process repeated until the column is
full. U-shaped columns are packed in the same manner. Columns up to 50 ft long
can be packed in a series of U's and then each U column joined with a low dead
volume connection. If the columns were glass they were usually filled through
an opening at the top of each U which was terminated in a plug of quartz wool
and sealed-off in a blow-pipe flame. These long packed columns could be
operated at a maximum of 200 psi. and could provide efficiencies of up to 50,00
theoretical plates. Such columns could tolerate charges of several microliters.
A chromatogram of the isomeric heptanes and octanes obtained from a 50 ft
YES GC-Columns Packing