Precision of n-Decyl beta-d Glucopyranoside by HPLC/ELSD

Application Notes

Precision of n-Decyl beta-d Glucopyranoside by HPLC/ELSD

The variation in the area counts of the peaks from the separations of the surfactant n-Decyl beta-d Glucopyranoside by reversed phase HPLC and evaporative light scattering detection was determined to be less than 1%RSD.

Reproducibility- n-Decyl beta-d Glucopyranoside, Alltech Application Note 0022E, May 2, 2000.

The variation in the area counts of the peaks from 11 separations of the surfactant n-Decyl beta-d Glucopyranoside by reversed phase HPLC and evaporative light scattering detection was determined to be less than 1%RSD.

Separation was achieved isocratically with methanol/water by reversed phase HPLC in around 3 minutes using an Alltech Econosphere C18, 5µm, 150 x 4.6mm column (Part No. 70065) and detection was by evaporative light scattering using an Alltech 500 ELSD (Evaporative Light Scattering Detector).

In ELSD, the mobile phase is first evaporated. Solid particles remaining from the sample are then carried in the form of a mist into a cell where they are detected by a laser.