Preparative Chromatography: Volume Overload


Preparative Chromatography
by RPW Scott
part of the Chrom Ed. Series


Free book pdf downloads

Similarly, in the chromatograms shown on the right hand side of figure 3, 2, 4 and 6 ml of a solution each containing 9.0 mg of naphthalene was injected onto the column and. The maximum sample volume that just permits the separation of naphthalene and anthracene, identified experimentally was 6 ml (theoretical value calculated from equation 2 was 6.1 ml). Destefano and Beachel (4) has also investigated the effect of volume and mass overload on resolution. They concluded that, given the choice, it is advantageous to overload a column with a large volume of a dilute solution of sample, as an alternative to using a small volume containing a high concentration of sample. They reported, however, that the validity of this conclusion, appeared to depend somewhat on the capacity ratios of the eluting solutes. In figure 4 the retention distance (measured in cm along the chart) is plotted against the sample volume

After, J. Chromatogr., Ref. [3]

 

Figure 4. Retention Distance of Benzene, Naphthalene and Anthracene against Sample Volume

 
 
 

Journal of Chromatographic Science