Figure 51 Optical Diagram of a Low Angle Laser Light Scattering Detector
Between the forward detector lens and the rear detector lens is A filter holder and an analyzer/polarizer. Finally the light is focused through a sensor aperture to an opal diffuser that spreads the scattered light through a red filter and onto the photo- multiplier.
The device is
frequently used with a refractive index detector in series to coincidentally
measure the refractive index of the eluent. This is necessary to calculate (K)
from the refractive index. As discussed, the molecular weight of a solute is
determined from the intercept of the graph relating (
)
to the solute concentration (c) as shown in figure 52. The concentration
calculated from the refractive index detector employing the data from prior
calibration.